FAQs

High conductivity

1. Muaj cov muaju-conductivity crystals nyob rau hauv conductivity cell.
Kev daws: Tom qab ntxuav lub cell conductivity nrog 1: 1 nitric acid, yaug nws nrog dej deionized.

2. Lub eluent tsis ntshiab txaus.
Kev daws: hloov qhov eluent.

3. Cov kab chromatographic absorbs high-conductivity tshuaj.
Kev daws: Ntxuav kab rov ua dua thiab sib hloov nrog eluent thiab dej.

4. Kev xaiv tsis raug ntawm kev ntsuas ntsuas
Thaum ua qhov zoo ions tsom xam, raws li eluate tus keeb kwm yav dhau conductivity siab dhau lawm, kev xaiv ntawm qhov ntsuas qis dhau yuav ua rau qhov qhia tau tias muaj txiaj ntsig zoo heev.Tsuas xaiv qhov ntsuas ntsuas dua.

5. Cov tshuaj tua kab mob tsis ua haujlwm
Kev daws: Txheeb xyuas seb lub tshuab hluav taws xob puas tau qhib.

6. Cov qauv concentration siab dhau.
Kev daws: Dilute tus qauv.

Siab hloov pauv

1. Muaj npuas nyob hauv lub twj tso kua mis.
Kev daws: Counterclockwise loosening pa valve ntawm twj tso kua mis valve, exhausting npuas.

2. Lub check valve ntawm lub twj tso kua mis muaj kuab paug los yog puas.
Kev daws: Hloov cov tshev valve lossis muab tso rau hauv 1: 1 nitric tov rau supersonic tu.

3. Cov lim dej hauv lub raj mis eluent muaj kab mob los yog thaiv.
Kev daws: Hloov lub lim.

4. Tsis txaus degassing ntawm eluent.
Kev daws: Hloov qhov eluent.

Lub 6-txoj kev txhaj tshuaj valve raug thaiv.

Kev daws: Tshawb xyuas qhov chaw clogging raws qhov kev taw qhia ntws los txheeb xyuas thiab tshem tawm qhov txhaws.

Nquag overpressure

1. Lub kem lim membrane yog thaiv.
Kev daws: Tshem cov kab thiab unscrew lub inlet kawg.Tshem tawm lub phaj kom zoo, muab tso rau hauv 1: 1 nitric acid thiab ntxuav nws nrog ultrasonic yoj rau 30 feeb, tom qab ntawd yaug nws nrog dej deionized thiab sib sau ua ke rov qab, rov qab sib sau cov chromatograph rau yaug.Nco ntsoov tias chromatograph tsis tuaj yeem txuas nrog txoj kev ntws.

2. Lub 6-txoj kev txhaj tshuaj valve yog thaiv.
Kev daws: Tshawb xyuas qhov chaw clogging raws li cov kev taw qhia ntws los txheeb xyuas thiab daws teeb meem.

3. Lub check valve ntawm lub twj tso kua mis yog thaiv.
Kev daws: Hloov cov tshev valve lossis muab tso rau hauv 1: 1 nitric tov rau supersonic tu.

4. Txoj kev khiav yog thaiv.
Kev daws: nrhiav qhov chaw clogging raws li maj mam tshem tawm cov txheej txheem thiab hloov pauv.

5. Tshaj tshaj tawm.
Kev daws: Kho lub twj tso kua mis kom haum tus nqi txaus.

6. Lub siab tshaj plaws txwv ntawm lub twj tso kua mis yog teem tsawg heev.
Kev daws: Nyob rau hauv kev khiav hauj lwm ntawm cov chromatographic kem, tswj lub siab tshaj plaws txwv siab mus rau 5 MPa saum toj no lub siab ua hauj lwm tam sim no.

Lub suab nrov nrov

1. Lub cuab yeej tsis ua haujlwm ntev txaus raws li tau npaj tseg.
Kev daws: Nruam infusion ntawm eluent kom txog thaum ntsuas ntsuas ruaj khov.

2. Muaj npuas hauv lub twj tso kua mis.
Kev daws: Counterclockwise loosening pa valve ntawm twj tso kua mis valve, exhausting npuas.

3. Lub lim dej ntawm cov yeeb nkab dej ntawm lub twj tso kua mis raug thaiv, ua rau lub siab tsis zoo nyob rau hauv lub zog nqus thiab tsim cov npuas.
Kev daws: Hloov cov lim lossis tso lub lim rau hauv 1: 1 1M nitric acid yuav tsum tau ntxuav 5min nrog ultrasonic da dej.

4. Muaj npuas nyob hauv kem.
Kev daws: Siv cov eluent npaj los ntawm cov dej deionized los yaug lub kem hauv qis ceev kom tshem tawm cov npuas.

5. Muaj npuas nyob hauv txoj kev ntws.
Kev daws: Tshem cov kab thiab cov npuas hauv dej.

6. Muaj npuas nyob rau hauv lub conductivity cell, uas ua rau tsis tu ncua hloov ntawm lub hauv paus.
Kev daws: Flushing conductivity cel, exhausting npuas

7. Qhov hluav taws xob tsis ruaj khov lossis cuam tshuam nrog hluav taws xob zoo li qub.
Kev daws: Ntxiv ib qho hluav taws xob stabilizer thiab av lub ntsuas.

High lub hauv paus hloov

1. Lub sij hawm ua ntej cua sov ntawm lub cuab yeej tsis txaus.
Kev daws: Txuas lub sijhawm ua ntej cua sov.

2. Dej ntws tawm.
Kev daws: nrhiav qhov chaw tawm thiab kho nws, yog tias nws tsis tuaj yeem daws tau, hloov qhov sib koom ua ke.

3. Qhov hluav taws xob tsis ruaj khov lossis cuam tshuam nrog hluav taws xob zoo li qub.
Kev daws: Ntxiv ib qho hluav taws xob stabilizer thiab av lub ntsuas.

Kev daws teeb meem tsawg

1. Lub concentration ntawm eluent yog tsis tsim nyog.
Kev daws: Xaiv cov concentration kom raug.

2. Tus nqi ntws ntawm eluentis siab dhau lawm.
Kev daws: Xaiv qhov txaus txaus tus nqi ntawm eluent.

3. Siv cov qauv uas muaj ntau ntau
Kev daws: Dilute tus qauv.

4. Kem muaj kab mob.
Kev daws: Rov tsim dua lossis hloov kab ke.

Tsis zoo repeatability

1. Lub ntim ntim ntawm cov qauv tsis tas li.
Kev daws: Txhaj cov qauv ntawm lub ntim ntau tshaj 10 npaug ntawm lub nplhaib kom muaj nuj nqis kom muaj kev txhaj tshuaj tag nrho.

2. Qhov concentration ntawm cov qauv txhaj tshuaj yog tsis tsim nyog.
Kev daws: Xaiv qhov concentration ntawm cov qauv txhaj tshuaj kom raug.

3. Cov reagent tsis huv.
Kev daws: Hloov cov reagent.

4. Cov khoom txawv teb chaws muaj nyob hauv cov dej deionized.
Kev daws: Hloov cov dej deionized.

5. Kev hloov pauv.
Kev daws: Tshawb nrhiav cov laj thawj ntawm cov kev hloov pauv no thiab kho nws mus rau qhov qub.

6. Txoj kev khiav yog thaiv.
Kev daws: nrhiav qhov chaw thaiv, kho lossis hloov pauv.

Redundant ncov

1. Cov reagent tsis huv.
Kev daws: Hloov cov reagents.

2. Deionized dej muaj impurities.
Kev daws: Hloov cov dej deionized.

Tsis Peak

1. Kev teeb tsa tsis raug ntawm cov khoom siv hluav taws xob.
Kev daws: Re-install lub conductivity cell.

2. conductivity conductivity cell puas lawm.
Kev daws: Hloov cov cell conductivity.

3. Lub twj tso kua mis tsis muaj cov tshuaj tso tawm.
Kev daws: Txheeb xyuas qhov ntsuas siab kom paub meej tias lub twj tso kua mis puas ua haujlwm.

Tsis zoo linearity

1. Cov txheej txheem kev daws teeb meem yog paug, tshwj xeeb tshaj yog cov qauv qis.
Kev daws: Rov npaj cov tshuaj.

2. Deionized dej tsis huv.
Kev daws: hloov cov dej deionized.

3. Cov concentration ntawm cov qauv yog siab dhau los yog tsawg dhau, tawm ntawm cov kab ntau ntawm cov khoom siv.
Kev daws: Xaiv qhov ntau ntawm cov concentration.

Qhov txawv txav tam sim no ntawm tus suppressor.

Kev daws: hloov cov hlua hluav taws xob lossis cov khoom siv hluav taws xob tas li.

Kev tsim cov npuas hauv lub twj tso kua mis

1. Absorbed roj nyob rau hauv txoj kev ntws cov yeeb nkab
Kev daws: Thaum cov dej tso rau, qhib lub qhov tso tawm ntawm lub twj tso kua mis, pib lub twj tso kua mis plunger thiab tas li vibrate lub lim kom tshem tawm tag nrho cov roj.

2. Kev kub siab sab hauv tsev ua rau tsis txaus degassing ntawm cov dej deionized.
Kev daws: Siv cov cuab yeej degassing hauv online.

3. Lub check valve ntawm lub twj tso kua mis muaj kuab paug los yog puas.
Kev daws: Hloov cov tshev valve lossis muab tso rau hauv 1: 1 nitric tov rau supersonic tu.